HANNA® instruments Catálogo General v39

Waterproof Quick Connect DIN Connector The meter connects to the multiparameter probe through a single waterproof connector and makes attaching and removing the probe quick and easy. The meter automatically detects the probe when connected. Data Transfer & Charging The USB Type-C port provides for easy data transfer to memory stick, PC, or other compatible devices and is used to charge internal lithiumion battery. Waterproof Probe Connection and USB Type-C Specifications HI98494 pH / mV Range 0.00 to 14.00 pH* / ±600.0 mV Resolution 0.01 pH / 0.1 mV Accuracy ±0.02 pH / ±0.5 mV Calibration automatic one-point Quick Calibration using HI9828-25; automatic one, two, or three points with automatic recognition of five standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer ORP Range ±2000.0 mV Resolution 0.1 mV Accuracy ±1.0 mV Calibration manual at one custom point (relative mV) Dissolved Oxygen Range 0.0 to 500.0%; 0.00 to 50.00 ppm (mg/L) Resolution 0.1%; 0.01 ppm (mg/L) Accuracy ±1.5% of reading ± 0.01mg/L for 0.00-20.00mg/L; ±5% of reading for 20.00-50.00mg/L; ±1.5% of reading ±0.1% for 0.0200.0%; ±5% of reading for 200.0-500.0% Calibration automatic Quick Calibration in water saturated air; one or two-point automatic calibration at 100% and 0%; manual single point using a value entered by the user in % saturation or mg/L Altitude Compensation, automatic: 420 to 850 mmHg *The range may be limited by the sensor’s limits. 7 Multiparameter 7.60 | www.hannainst.com portable